Operation
5.2
Every measuring channel must be calibrated for the installation environment and the target prior to measur-
ing. The two potentiometers are set for calibration.
- Turn the potentiometer in a clockwise direction to effect a positive change in the output voltage.
- Turn the potentiometer in counterclockwise direction to effect a negative change in the output voltage.
- Use the original measuring environment for calibration where possible. Simulate the measuring environ-
Step 1: Zero adjustment
+5 V
-5 V
Fig. 15 Output signal on the controller, optimum zero point
NLS3181

Calibration

ment if this is not possible.
The measuring system operates. Connect the analog output with an oscilloscope.
Shift the zero point with the potentiometer "zero" until you get a symmetric output signal, see
0 V
Zero point
shifting
Fig.
15.