Operation
5.1
The controller contains two potentiometers, see Fig. 13, for adjusting zero and gain. The measuring signal
can be adjusted in the range from -5 VDC to +5 VDC at a max. output current of 10 mA. The output is tempo-
rarily short-circuit-proof. The output impedance is less than 10 ohms.
i
Fig. 14 Potentiometer for calibration
NLS3181
Control Elements
The potentiometers Zero and Gain have 24 turns! A slight click is audible at the end positions.
Gain
Zero point
Page 51