GUIDANCE AND MANUFACTURER'S DECLARATION- ELECTROMAGNETIC IMMUNITY:

This device is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Basic EMC
standard
Electrostatic
Discharge (ESD)
IEC61000-4-2
Electrical fast
transient/ burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage
dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
Conducted RF
IEC 61000-4-6
Voltar ao índice
20
Immunity Test Levels
HOME HEALTHCARE
ENVIRONMENT
±8kV contact
±15kV air
±2kV for power supply
line
±1kV for input/output line
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
Voltage Dips:
i) 100% reduction for 0.5
period,
ii) 100% reduction for 1
period,
iii) 30% reduction for
25/30 period,
Voltage Interruptions:
100% reduction for
250/300 period
30 A/m
3 Vrms
0,15 MHz – 80 MHz
6 Vrms in ISM and
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
(4)
Compliance Levels
±8kV contact
±15kV air
±2kV for power
supply line
±1kV for input/output
line
± 1 kV line(s) to
line(s)
230V (U
)
(1)
T
Voltage Dips:
i) 100% reduction for
0.5 period,
ii) 100% reduction for
1 period,
iii) 30% reduction for
25/30 period,
Voltage Interruptions:
100% reduction for
250/300 period
30 A/m
6Vrms
836001-5210 V2.5
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions,
it is recommended that the device
be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
O equipamento de comunicação
portátil e RF móvel não deve ser
usado próximo de qualquer parte
deste dispositivo, incluindo os
cabos, a uma distância inferior à
recomendada e calculada segundo
a equação aplicável à frequência do
transmissor.
www.stryker.com