Use the
overlapped images between live ultrasound and the registered slice of the
3D volume from the 2nd modality.
Fine Registration - With One Point
1. Click on
One Point
modality 3D volume.
2. Find an anatomical point on the live ultrasound (left side)
3. Click on
Ultrasound
image
4. Click on the anatomical point to be registered on the ultrasound image
5. Move the probe to find the 2nd modality slice with the same
anatomical point
6. Click on
2nd Modality
7. Click on the anatomical point to be registered on the 2nd modality
The live Ultrasound image is now perfectly synchronized with the 2nd
Modality on a specific anatomical point.
When you want to register more than one internal marker, go to the
Internal Markers Registration section.
TouchRing
Use the
ultrasound image and the 2nd modality (Fusion blending control).
WARNING
After registration and particularly before using the system for the current
session, it is mandatory to check the system registration accuracy by
performing a side-by-side comparison of the ultrasound live image and
the registered 2nd modality image using the TouchRing or PW Overlap
controls, to determine how well the two images match.
If the ultrasound live image and the corresponding 2nd modality image
appear to be poorly registered, repeat the registration procedure.
Aixplorer® Fusion/Navigation User's Guide
PW
or
knob to change the priority on the
to better match the live ultrasound with the 2nd
under Record Position to freeze the ultrasound
under Record Position to freeze the image
PW
or
to change the priority between the
83