Cleaning and Disinfecting the
eTRAX™
1. After each use, remove needle sensor from needle, discarding single-
use, disposable components. Remove visible contaminants from
needle sensor surface.
2. Clean and disinfect all surfaces with common germicidal or antiseptic
such as Sani-cloth® HB, CIDEX OPA® (Johnson & Johnson),
ENZOL® (Johnson & Johnson), T-Spray™, Protex™ or alcohol.
3. Follow manufacturer's directions for all cleaning and disinfection.
Cleaning and Disinfecting the
General Purpose Sensor for
the omniTRAX™ CT and MR
and VirtuTRAX™
This section explains how to clean, disinfect and sterilize the general
purpose sensor used for the VirtuTRAX and omniTRAX devices. The
other accessories used for these devices are single-use, and therefore
disposable.
WARNING
Users of this product have an obligation and responsibility to provide
the highest degree of infection control to patients, co-workers and
themselves. To avoid cross-contamination, follow infection control
policies established by your facility.
The recommended protocol for reprocessing the general purpose sensor
has been evaluated for effectiveness and compatibility. If choosing to use
a reprocessing protocol other than those listed, a validation must first be
completed to ensure effectiveness and compatibility.
Do not gas sterilize or autoclave general purpose sensor.