Cleaning, Disinfecting and
Sterilizing the Bracket

Cleaning the Bracket

Reusable devices must be thoroughly cleaned before they are subjected
to the sterilization or high-level disinfection. Brushes should be used
to thoroughly clean all surfaces. Use enzymatic cleaners for removing
proteins from reusable devices.
Remember, you cannot achieve sterilization or high-level disinfection
unless the assembly is cleaned first.
When cleaning and sterilizing reusable devices, be sure to:
• Thoroughly clean all surfaces of reusable components with a brush
while rinsing to loosen and remove materials.
• Thoroughly examine all surfaces, visually inspect the ENTIRE device
is clean and no visible things remain.
• Flush with the high-level disinfectant to ensure that the disinfectant
reaches all areas.
• Always use sterile water for rinsing or removing residual germicides
from devices. Do not rinse reprocessed devices with tap water, which
may recontaminate the device.
• Thoroughly dry the device after rinsing.
• Always use a probe cover to minimize any contaminants during the
bracket usage.
• After sterilization or high-level disinfection, appropriately package
and store the device to ensure that sterility is not compromised prior
to reuse.
High-level Disinfecting and
Sterilizing the Bracket
After cleaning by using the above recommendations, soak for a minimum
of 5minutes in a neutral pH Enzol Enzymatic detergent such as made by
Johnson & Johnson©.
Aixplorer® Fusion/Navigation User's Guide