Symptom
Data is too noisy
Poor accuracy
110
Aixplorer® Fusion/Navigation User's Guide
Possible Causes
Filters OFF
Low signal
External noise
in environment
Changing hemisphere
Line frequency
value set incorrectly
Metal in tracking
environment
Damaged equipment
Sensor or transmitter
connector not
properly inserted
A software application error
Board error
Solution
- Check that sensor/
transmitter connectors are
correctly installed. Inspect
pins for wear or damage.
- Contact technical
support for assistance
- Check FILTER
STATUS
- Decrease distance from
sensor to transmitter
Be sure that the sensor
is not located near the
tracker's power supply
or other electronic
devices or cables.
If the signs of the X, Y
or Z position outputs
suddenly change, you
may have crossed a
hemisphere boundary.
Use the HEMISPHERE
command to rectify
Determine correct
frequency (50 Hz
in Europe, 60 Hz in
North America) and
set to correct value
Check all around the
transmitter to the furthest
distance from the centre
of the transmitter to
the maximum distance
the sensor is used.
Move or replace
metal, or reposition
the driveBAY system
Check for damage
Correctly by unplugging
and plugging the
components back
into the board
Verify formulas
and scale factors
Board error