If you encounter difficulty in the operation of the Fusion feature, use the
information in this section to help correct the problem.
If the problem is not covered here, contact your Customer Support
Representative.
See
the section called "driveBAY" [22]
status.
The troubleshooting table below contains a list of symptoms and the
action to take to correct the problem.
Symptom
No front panel
LED illumination
Front panel LED
blinking red/yellow
Demo Utility doesn't run
Power-up defaults
did not change after
configuring with the utility
Not able to
communicate with
the system using USB
Can communicate
with the system, but
data not changing
Aixplorer® Fusion/Navigation User's Guide
for more details about LED
Possible Causes
No power
No valid transmitter attached
No serial communication
Configuration
download interrupted
driveBAY did not reset
(restart) after burning
the Flash settings
Re-initialize USB
Sensor is saturated
Transmitter is OFF
and disconnected
Component
connections faulty
Solution
- Check AC connections
to power supply
- Reset hardware by
cycling AC power
- Attach transmitter
- Reset transmitter
connector
- Check the suggestions
outlined in 'Not able to
communicate' below
- Re-start the driveBAY
and the utility and set the
defaults again. Be sure to
click APPLY to send the
setting to the Electronics
- Cycle power to the
driveBAY, and re-
check the settings
- MDSP failed to
respond to messages
- Cycle power
on the driveBAY
- Check the error codes
for a sensor a saturation
condition. Move the
sensor farther away
from the transmitter.
- Check status of the
transmitter. Turn on using
SELECT_TRANSMITTER
parameter (3D
Guidance API)
109