2. GENERAL DESCRIPTION

2.1 FRONT VIEW

Hole for
emergency
Union joint, to
Union joint, to
fill the tank with
let out the waste
distilled water
water

2.3 REAR VIEW

Union joint, to let out
3. INSTALLATION / START UP

3.1. INSTALLATION

After removing the autoclave from the packaging, it should be installed as follows:
• The autoclave should be placed on a flat levelled surface, leaving an emptyspace of a 4 cm
at the back of the machine with a width of at least :
• Quaz 17 l : 50 cm
• Quaz 24 l : 65 cm
It can also be installed on a smaller surface, moving the front legs from position 1 to position 2
(or 3 only for Quaz 24 l).
• Quaz 17 l : Position 2, 45,5 cm
• Quaz 24 l : Position 2, 59 cm, Position 3, 55 cm
We recommend that the legs are kept in their original position (the most external one) to
make the machine more stable.
We recommend leaving an empty space of at least 3 cm at the top and the sides of the
machine, so it has enough ventilation.
• Do not install the autoclave in a damp place.
• Do not place any objects on it.
• Keep it away from any sources of heat such as radiators or windows where there is direct
sunlight.
• Install it in a well ventilated place.
• Keep the machine away from places where there is a risk of explosion.
• Install the machine on a levelled surface.
IMPORTANT:
Do not tip the machine when installing it, it may get damaged inside. Always keep it in
vertical position.
WARNING:
• The door will be blocked until the autoclave is connected to the power supply.
• The machine has a slight inclination at the back to facilitate the draining of the
sterilisation vessel. In case the surface where it is installed is not levelled, this
inclination should be maintained by adjusting the front legs.
8
Touch-screen
Bacteriological
Front
filter
legs
Refrigeration grid
RS 232 connector
to external
'QUAZ Printer'
USB Connector
the distilled water
Rear wheels

2.2 FRONT VIEW OPEN DOOR

Latch
Handle to open
the door
Door
Main switch (side)
Memory card
slot + top led
Silicone gasket
Conductivimeter plug
(only in 5.17 and 5.24 versions)
Safety valve
Power cord
17 l
• Always keep the door closed when moving the machine, do not use it as a support.
• Do not put pressure on the edge of the door while it is open (especially when the front
legs are in the second position).
Hinge
Stainless steel
Sterilisation chamber
cover
Tank air inlet
Connection for the automatic
filling of distilled water
QUAZ label
Connection for the automatic
emptying of waste water
24 l
Pos. 2
Pos. 1
Support for
trays and
sterilisation
vessels
Sterilisation
trays
Pos. 3
Pos. 2
Pos. 1