Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance; Regular Maintenance - Satelec ACTEON Quaz Class B Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 65
Excessive temperature.
ERROR 005
- The temperature inside the sterilisation chamber has exceeded the established
(050,051,052
limits for temperature in the stage in process.
053,054)
- The temperature inside has exceeded the established limit.
- The temperature of the heating devices has exceeded the established limit.
ERROR 006
Vacuum pump failure.
(060,061,062,
During the Vacuum Test the vacuum pump has not reached the established
063,064,065)
vacuum values, the fractionated vacuum or drying, within the established time.
ERROR 007
Faulty sterilisation.
(070,071,072,
During the sterilisation stage the temperature or pressure values are different
073,074,075,
to the established values and to the values accepted by standard
076,077,078)
EN 13060:2004+A1:2009.
Faulty decompression.
ERROR 008
If the machine does not reach the values of decompression within the establi-
(080,081,082)
shed time.
Faulty ventilation.
ERROR 009
If the pressure has not been restored during the ventilation stage within the
established time.
Power failure.
ERROR 010
If there is a power cut during a cycle, because the supply fails or because the
machine's own safety devices fail.
Interrupted cycle.
ERROR 011
The user cancels the sterilisation cycle deliberately.
ERROR 012
Temperature system failure.
(120, 121, 122
- When the system detects that the cable of a temperature sensor is damaged.
123,124)
- When the system detects a malfunction of the room temperature sensor inside.
Pressure system failure.
ERROR 013
When the system detects that the cable of the pressure sensor is damaged.
Lack of leak tightness.
When the loss values are higher than the values required by standard EN
ERROR 014
13060:2004+A1:2009. (10% of P0 - P1 for the stabilisation time or 10% P0 - P2 for
the measurement time.)
ERROR 015
Power supply failure.
(150,151,152)
There is not enough voltage in the internal power supply of the machine.
ERROR 016
Door malfunction.
(160)
The electrical security locks on the door detect an incorrect position.

6. MAINTENANCE

The QUAZ autoclave, like all autoclaves, should not only be used properly, but it also
requires very special periodic control and maintenance.
This ensures that the autoclave will operate safely and efficiently.
The controls required include:
- Regular maintenance that can be carried out directly by the user.
- Preventive maintenance that should be carried out by the Authorised Technical Support.
IMPORTANT: In case any pieces need replacing, only use original parts.

6.1. REGULAR MAINTENANCE

IMPORTANT: Before performing any kind of maintenance, check that: The autoclave is
disconnected from the power supply, that is it not hot and that there is no pressure
inside the sterilisation vessel.
- For the machine to stay in good condition, periodically clean all the external parts using
a damp cloth with common neutral detergent (do not use abrasive or corrosive products).
- Frequently check and clean the leak tightness silicone gasket, the door disk and the edge
of the sterilisation vessel with a damp cloth (do not use alcohol).
- Do not use abrasive cloths or (abrasive) metallic brushes to clean metal parts.
18
DESCRIPTION
Allow the autoclave to cool down and repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
Make sure that there is nothing obstructing the ventilation system.
Make sure that the fans are in good working order
Allow the autoclave to cool down and repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
Allow the autoclave to cool down and repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
Allow the autoclave to cool down and repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
Check the state of the bacteriological filter.
If the problem persists, please contact the Technical Support Dept.
Check that there is a constant power supply.
Allow the autoclave to cool down and repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
Switch off and on back again the sterilizer. If problem persists, please contact
the Technical Support Dept.
Switch off and on back again the sterilizer. If problem persists, please contact
the Technical Support Dept.
Check the state of the silicone gasket and make sure the door is closed properly.
Repeat the cycle.
If the problem persists, please contact the Technical Support Dept.
Check that there is a constant power supply.
If the problem persists, please contact the Technical Support Dept.
Follow the instructions on the screen.
Turn the emergency key in any direction until it stops (approx 25 times)
If the problem persists, please contact the Technical Support Dept.
- In case the autoclave is not going to be used for a period of time, empty the waste water
tank and distilled water tank completely (Use the corresponding tubes suppliedwith the
machine).
The following Regular Maintenance Schedule that shows how often the maintenance should
be carried out:
ACTION
---
Union joint to empty
The distilled water
Tank completely

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis