4.3.6.4 USERNAME

4.3.6.5 MAINTENANCE: COUNTER / CYCLE RESET

This screen displays the cycles that are left to:
- Replace the bacteriological filter
- Change the silicone gasket on the door
- Require maintenance by a specialised engineer
- "Water Quality" is about the quality of the water measured by the optional water
quality sensor

4.3.6.6 CYCLES COMPLETED AND TIME OF OPERATION

On this screen the user can see the statistics of the cycles that have been completed
(number of cycles completed of each type and average of cycles per day) as well as the
time of operation in relation to the time the machine has been connected to the power
supply.
14
Exit screen
Select the
See the selected
option
option
Exit screen
Select the
option
See the selected
Exit screen and save the
option
The autoclave gives a warning 50 cycles before maintenance is required. This message will
be repeated every 10 cycles, only once a day, until maintenance is carried out.
Once the required maintenance has been carried out the counter will return to the
original number (500 or 1,000) when it is selected with the cursor and the RESET icon is
pressed. The autoclave saves the date of each reset after maintenance.
4.4. TEST CYCLES (VACUUM AND BOWIE & DICK)
On this screen the test cycles Vacuum and Bowie & Dick can be selected.

4.4.1. VACUUM TEST

If the door is closed and the VACUUM TEST icon is pressed, the 'VACUUM TEST' cycle will start
automatically.
IMPORTANT:
- The VACUUM TEST performs a leak tightness test (to check that there are no
microleaks) in the hydraulic circuit.
- The VACUUM TEST should be carried out when the autoclave is at room temperature.
- The VACUUM TEST should be carried out with the sterilisation chamber empty.
- The values of the VACUUM TEST are expressed in absolute pressure values.
Preheating on
Preheating off
Exit screen and save the
selected option
In this option the user has 40 characters available. The
information will be saved in the autoclave identification
screen (see section 4.1) and in the sterilisation record.
Select alphanumeric
character
Confirm alphanumeric
character
information
There are 3 possible cases in the "Water Quality" menu:
- Empty. The autoclave does not have associated a water
quality sensor. If the autoclave has one, then it hasn't
detected yet a non acceptable quality water
- OK: The water quality is OK
- OK: The water quality is not OK. If there is a deminer
lizator unit, change the cartridges.