Herunterladen Inhalt Inhalt Diese Seite drucken

Scope Of Delivery - Mettler Toledo InPro 6850 i Bedienungsanleitung

O2
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

InPro 6850 i / 6900 i / 6950 i O
ISM Sensors:
All oxygen sensors with the index "i" (6850 i, 6900 i, 
6950 i) are equipped with ISM.
Principle: In the sensor head a chip is integrated, 
which takes over the entire monitoring and control of 
the sensor and, beyond that, stores all sensor data. 
This chip is responsed via the transmitter.
The following data are available permanently in the 
sensor:
–  type of sensor
–  serial no.
–  software version
–  hardware version
–  order no.
–  operating time
–  calibration time and calibration date
–  calibration table
To check the system, the following indicators are 
supervised:
–  temperature
–  slope
–  zero current
–  air current
–  polarization voltage
Based on these informations, the transmitter cal-
culates a wear monitor and displays it depending 
upon transmitter type differently. (see the respective 
instruction manuals) 
ISM enables the connection of the digital sensor to 
the iSense™ Asset Suite. The user-friendly software 
allows to manage all sensor information and store 
it in a data base. Furthermore, the sensor can be 
calibrated via the iSense™ software.
3.3

Scope of delivery

Each sensor is supplied fully assembled and factory-
tested for correct function together with: 
–  an electrolyte bottle (52 206 111)
–  a quality control certificate
–  inspection certificates 3.1
(complying with EN 10204)
Digital sensors must be filled with electrolyte before 
start-up.
© 04 / 13 Mettler-Toledo AG   
Printed in Switzerland 
 Sensor 12 / 25 mm  
2
InPro 6850 i / 6900 i / 6950 i 
13
52 206 349

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Inpro 6900 iInpro 6950 i