InPro 6850 i / 6900 i / 6950 i O
2.6
Ex-classification ATEX (not for InPro 6950 i)
2.6.1 Introduction
According to RL 94/9/EG (ATEX 95) Appendix l,
lnPro 6XXX*/*/*/*/* oxygen sensors are devices
group ll, category 1/2G and according to RL 99/92/
EG (ATEX 137) may be used in zones 0/1 or 0/2
and gas groups llA, llB and llC that are potentially
explosive due to c ombustible substances in the
temperatures classes T3 to T6.
For use/installation, the requirements of EN 60079-
14 must be observed.
According to RL 94/9/EG (ATEX 95) Appendix l,
lnPro 6XXX*/*/*/*/* oxygen sensors are devices
group ll, category 1/2D and according to RL 99/92/
EG (ATEX 137) may also be used in zones 20/21
that contain combustible dusts.
For use / installation, the requirements of EN 61241-
14 must be observed.
The digital sensor circuit is part of a common intrin-
sically safe system and is for opera t ion connected to
a separately certified transmitter.
The digital sensor circuit as part of an i ntrinsically
safe system is isolated from the not- i ntrinsically safe
electric circuits up to a maximum r ated voltage of
375 V and from grounded parts up to a maximum
rated voltage of 30 V.
2.6.2 Rated data
Measuring circuit:
Fail-safe ignition protection class Ex ia IIC only for
connection to a certified fail-safe circuit.
Maximum values:
U
# 16 V, I
i
L
= negligible
i
C
= negligible
i
Note:
The above maximum values are each the total of all
i ndividual circuits of the associated intrinsically safe
power supply and transmitter.