Maintenance; Inspection Of The Sensor; Visual Inspection - Mettler Toledo InPro 6850 i Bedienungsanleitung

O2
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

20 
6

Maintenance

6.1

Inspection of the sensor

6.1.1 Visual inspection

To check your sensor, we recommend the following 
procedure:
–  The contacts of the connector must be dry. 
  M oisture, corrosion and dirt in the connector can 
lead to false readings.
–  Check the cable for buckling, brittle areas or 
  r uptures.
–  Before calibration always examine the membrane 
foil optically for signs of damage. The foil must 
be intact and clean. Dirty membranes should be 
  w ip  e d clean using a soft, moist tissue.
h
Note: An undulated membrane has no influence on
the sensor performance, assuming the membrane
is intact.
–  The membrane body must be replaced if the 
  s ensor has too long a response time, the reading 
is   u nstable or subject to drift, and if the sensor 
  c annot be calibrated or the membrane shows 
sign of   m echanical damage.
–  Check the cathode area for discoloration, conta-
mination or cracks in the glass. If necessary rinse 
with demineralized water and clean with a clean 
soft brush or soft paper tissue.
a
Attention! Do not use any cleaning agents contai-
ning alcohol. This could damage the sensor or
lead to fault current.
a
Attention! The glass body is fragile and sensitive
to vibration.
InPro 6850 i / 6900 i / 6950 i 
52 206 349 
InPro 6850 i / 6900 i / 6950 i O
 Sensor 12 / 25 mm
2
© 04 / 13 Mettler-Toledo AG
Printed in Switzerland