Product description

3.1

General information

The oxygen sensor series InPro 6850 i / 6900 i /
6950 i with integrated temperature probe are used 
for meas  u re  m ent of oxygen at low and medium 
  c oncentrations.
The sensors are sterilizable, and in the majority 
of cases autoclavable too and they are compatible
with CIP (Cleaning-In-Place).
InPro 6xxx i sensors with ISM functionality offer 
Plug and Measure as well as enhanced diagnostics 
  f eatures.
3.2

Principle

Amperometric oxygen sensors:
The amperometric oxygen sensors of METTLER  
TOLEDO base all on the same mearuring principle 
according to Clark. However different series are of-
fered, which clearly differ in the number and arrange-
ment of their electrodes and thus in their specifica-
tions:
•  The measurement system of the InPro sensors 
consists of a working electrode (cathode), a 
counter electrode (anode), a reference electrode 
and a guard ring electrode. The measurement 
system is separated from the process medium by 
an oxygen permeable membrane.
•  The measuring technique in principle is alike with 
all sensors. 
•  The transmitter supplies a constant voltage be-
tween cathode and anode.
•  The electrolyte creates a conductive connection 
between the electrodes.
•  The oxygen molecules migrate from the mea-
surement medium through the membrane to the 
cathode to which the voltage is supplied and are 
reduced. At the same time, oxidation takes place 
at the anode. 
•  This causes a current to flow between the anode 
and cathode which is directly proportional to the 
partial pressure of oxygen (pO
medium. With digital ISM sensors, the sensor 
itself converts the current into oxygen concentra-
tion and communicates the value to the transmit-
ter.
•  The guard ring electrode of the InPro 6900 i and 
6950 i built into the sensor reduces the oxygen 
that migrates to the side of the cathode and can 
distort the measurement. The guard ring electrode 
therefore enables traces of oxygen to be precisely 
determined at even the lowest concentrations.
h
Note: Please refer to "Section 10 – Theory of the
polarographic sensor" for further information.
InPro 6850 i / 6900 i / 6950 i 