8.2

Specifications

   
Measurement principle   
Working conditions
  Pressure resistance  
  (measurement)  
   
   
  Mechanical pressure resistance  Max. 12 bar
  Temperature range  
  (measurement)  
  Temperature range  
  (mechanical)   
   
   
   
   
   
   
Construction
  Temperature compensation 
  Cable connection 
  O-ring material   
   
  Membrane material 
   
  Material sensor body  
  (wetted parts) 
  Surface roughness of     
  wetted metal parts
  Quick disconnect interior body 
  Cathode 
  Anode   
   
   
  Guard ring 
   
   
  Reference 
Dimensions
  Sensor diameter  
   
   
  Immersion length (a)  
  for 12 mm sensor 
   
   
   
  Immersion length (a)  
  for 25 mm sensor 
   
   
Performance
  Detection limit   
   
   
  Accuracy 
   
   
  Response time    
  at 25 °C / 77 °F
  Sensor signal in ambient air  
  at 25 °C / 77 °F   
   
InPro 6850 i / 6900 i / 6950 i 
52 206 349 
InPro 6850 i / 6900 i / 6950 i O
InPro 6850 i / 6900 i / 6950 i
Polarographic Clark electrode
6850 i:  0.2 ... 6 bar  
6900 i:  0.2 ... 6 bar  
(9 bar with T-6900 R) 
6950 i:  0.2 ... 6 bar
0 ... 80 °C  
[32 ... 176 °F]
6850 i:   – 5 ... 140 °C    
[23 ... 284 °F] 
6900 i:   – 5 ... 121 °C 
[23 ... 249.8 °F] 
(sterilizable and autoclavable) 
6950 i:   – 5 ... 121 °C 
[23 ... 249.8 °F] 
(sterilizable)
Automatic 
AK9 
Silicone FDA and 
USP Class VI approved
PTFE / Silicone / PTFE 
(reinforced with steel mesh)
316L stainless steel  
N5 (Ra = 0.4 µm [16 µin]) 
Standard
Pt
6850 i:   Pt 
6900 i:   Ag 
6950 i:   Pt
6850 i:  – 
6900 i:  Pt 
6950 i:   Pt
Ag
6850 i:  12 / 25 mm 
6900 i:  12 mm 
6950 i:  12 mm
6850 i / 6900 i: 
70, 120, 220, 320, 420 mm   
[2.8, 4.7, 8.66, 12.6, 16.5 
6950 i:  70, 120, 220, 320 mm 
[2.8, 4.7, 8.66, 12.6 
6850 i:  80, 160, 260, 360 mm 
[3.2, 6.3, 10.2, 14.2 
6900 i:  – 
6950 i:  –
6850 i:  6 ppb to saturation 
6900 i:  1 ppb to saturation 
6950 i:  0.1 ppb to saturation
6850 i:  # ± [1 % + 6 ppb] 
6900 i:  # ± [1 % + 1 ppb]
6950 i:  # ± [1 % + 0.1 ppb]
98 % of final value in < 90 s 
6850 i:  50 ... 110 nA 
6900 i:  250 ... 500 nA 
6950 i:  2500 ... 6000 nA