K-Factor - bürkert 8045 Bedienungsanleitung

Magnetisch-induktiver durchflsstransmitter
Vorschau ausblenden Andere Handbücher für 8045:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4.4.3 K-Factor

The menu makes it possible to determine the K factor of the process:
either by entering the value of the K factor of the fitting (depending on the pipe
A
diameter and material, see instruction manual of the fitting S020): the transmitter
then calculates the K factor for the process using the fitting K factor, the cell
constant Fs and the temperature coefficient Kw.
Fs and Kw are set ex works and can be read and modified within the TEST menu.
or by using the „TEACH-IN" procedure. It consists either of a volume measurement
B
or a comparison measurement with another flow meter.
The value of the K factor taken into account for the process is the last determined K
factor, whatever the method used.
ENTER
K FACTOR
K= 001.00
Current process K
factor
CURRENT
CURRENT
CURRENT
*
0....9
Press simultaneously the keys to
shift the decimal point.
CURRENT
8045
CALIBRATION
ENTER
ENTER
K-FITING
A
0....9
ENTER
TEACH Y
B
ENTER
K= 001.68
Display of the
calculated K factor
ENTER
K= 001.68
Display of the calculated K factor
23
ENTER
K1= 001.00
*
0....9
Enter the K factor
of the fitting
(max. 999,99).
ENTER
ENTER
VOLUME
Start of
measurement
ENTER
End of
measurement
0....9
Enter the measured volume.
Same unit as for flow.
Start of
measurement
ENTER
FLOW
ENTER
ENTER
VALID N
The process K
factor is not re-
0....9
calculated.
VALID Y
ENTER
K= 001.69
FILL END
ENTER
0100.0 L
0....9
MEASURE
50 s.
measurement
ENTER
10.10 L/S
Enter the
measured flow
0....9