[STATus]: OPER:COND?
Description: This query command returns the content of Operational Condition
[STATus]: QUES:EVEN?
Description: This query command returns the content of Questionable Data Event
[STATus]: QUES:ENAB? <DNPD>
Description: This command enables bits in the Questionable Data Enable register.
Parameters: <DNPD> Represents the register bits as a decimal number.
Example:
Query:
[STATus]: QUES:COND?
Description: This query command returns the content of Questionable Condition
[STATus]: PRES
Description: This query command clears all bits in the Operation Data Enable
Using IEEE 488.2 Common Commands
The following commands are common to the IEEE 488.2 Standard. A summary of the
Calibrator's IEEE Common Commands are listed in Table XIV.
*IDN?
Description: This query command returns the manufacturer's name, model number,
The reply is formatted as follows: ORBIT CONTROLS,OCM191,xxxxxx,1.00
Where xxxxxx is serial number.
*OPC
Description: This command sets the OPC bit in the ESR (Event Status Register)
*OPC?
Description: This command returns "1" to the output queue after all pending
Betriebsanleitung
register. It's decimal value corresponds to the binary-weighted sum of
all bits set in the register. Register is not cleared after this query. The
response to the query therefore represents an instantaneous Snapshot
of the register state, at the time that the query was accepted.
register. It is a decimal value which corresponds to the binaryweighted
sum of all bits set in the register. Register is cleared after this query.
Selected bits are summarized at bit 3 (OSS) of the IEEE 488.2 Status
Byte register.
STAT:QUES:ENAB 2 Sets bit 1 of the Questionable Data Enable
SAF:QUES:ENAB?
register. It is a decimal value which corresponds to the binaryweighted
sum of all bits set in the register. The register is not cleared
after this query. The response to the query therefore represents an
instantaneous Snapshot of the register state, at the time that the query
was accepted.
register and in the Questionable Data Enable register.
serial number, and firmware revision of the Calibrator.
when all pending operations are complete.
operations are complete.
OCM191 Isolation Tester - Kalibrator
register, while all other bits are set to 0.
Returns the value of the Questionable Data
Enable register as a decimal value.