Herunterladen Inhalt Inhalt Diese Seite drucken

orbit controls OCM191 Betriebsanleitung Seite 42

Inhaltsverzeichnis

Werbung

Orbit Controls AG
[SOURce] :HVC[:LEV](?) <CPD> { 0 | 1 | 2 }
Description: This command selects one of three discrete capacitance closest to the
Parameters: <CPD>
Example:
Query:
[SOURce] : HVC:VOLT?
Description: This command returns the measured voltage on the output terminals.
Query:
[SOURce] : PSP:TOT?
Description: This command returns measure of total time.
Query:
[SOURce] : PSP:VOLT?
Description: This command returns the measured voltage on the output terminals.
Query:
[SOURce] : PSP:RES0(?) <DNPD>
Description: This command sets value of R0.
Parameters: <DNPD>
Example:
Query:
analogy for R1, R2, R3
[SOURce] : PSP:TTIM(?) <DNPD>
Description: This command sets value of T1.
Parameters: <DNPD>
Example:
Query:
analogy for T2, T3
[SOURce] : DPP:[LEV](?) <CPD>
Description: This command sets parameter in DPP mode.
Parameters: <CPD> =
Example:
Query:
Betriebsanleitung
Value specified by DNPD.
= 0 sets C0 capacitor
1 sets C1 capacitor
2 sets C2 capacitor
HVC C1
High voltage capacitance with 10nF.
HVC?
Returns the name of the selected resistance.
HVC:VOLT?
PSP:TOT?
PSP:VOLT?
PSP:RES0 1.25E+7 R0 in PSP mode is set to 12.5MΩ
PSP:RES0?
PSP:TTIM1 56
PSP: TTIM1?
0 is DAR parameter
1 is PI parameter
2 is PR parameter
DPP 1
Parameter is set to PI in DPP mode.
DPP?
Returns the name of the selected parameter.
OCM191 Isolation Tester - Kalibrator
Resistance value in ohms.
Returns the value of the R0.
Time value in seconds.
T1 in PSP mode is set to 56seconds
Returns the value of the T1.
42