12 Foot switch (accessory)
The following foot switch is offered for the CURIS
REF 36 01 10 → two-pedal foot switch without flushing pump button, explosion protected, 4 m
cable
REF 36 01 14 → two-pedal foot switch without flushing pump button, explosion protected, 4 m
cable
Intended use
The foot switch serves as a triggering mechanism of the CURIS
power output of the RF generator on and off.
REF 36 01 10
Maintenance, cleaning, disinfection
When these instructions are followed, the foot switch requires minimal maintenance. Depending
on the ambient conditions and frequency of use, regular maintenance and the inspection of the
housing and connecting cables for damage and problematic dirt is recommended.
For only manual cleaning, use just a cloth impregnated with water and a mild detergent. Do not use
cleaning agents which may damage the plastic surfaces such as detergents, abrasive cleansers, or
solvent-based cleaners.
NOTE
For safety reasons, always perform a functional test prior to surgical application. Always
activate the pedals with the electrosurgery device turned on. To prevent unintentional
burns, perform the functional test without electrode cables connected to the
electrosurgery device.
Technical data
Standards: IEC 60601-1:2005, IEC 60601-2-2:2009, IEC 60529:1989, IEC 60721-3-2:1997;
Class: Class I according to the Council Regulation (EU) 2017/745
Pedals made of break-proof, self-extinguishing thermoplastic material, cast aluminum housing
Connecting cable: permanently connected and sealed control cable
Protection class: IP X8 (1 m / 35 min.) according to IEC 60529
Switching element: reed contact
Switching voltage: max. 25 V AC / 60 V DC
Switching current: max. 1 A
Switching power max. 30 VA; mech. service life: >1 million switching cycles
Approvals: AP compatible
Storage conditions: Temperature: -25°C to +70°C; relative humidity: 5 % - 100 %; air pressure:
500 hPa – 1100 hPa
40
Yellow pedal → cutting mode activation
Blue pedal → coagulation mode activation
:
®
RF generator. It is used to turn the
®
CURIS
Instructions for Use
®
REF 36 01 14