CURIS
Instructions for Use
®
11 Error diagnosis
Error
Elements on the front stay
dark, device does not
function
Elements on the front work
normally, no RF output
power
Auto-Start button is
flashing
Display of an error
message, e.g. Err 42
Error messages and their meaning
The CURIS
RF generator conducts extensive functional tests during the power-on self-test.
®
All safety-related functions are also checked continuously during operation. In case of errors, the
RF energy output is stopped and the error is indicated by an error message.
NOTE
After an error message is displayed, the device can only be reset by turning it off and
back on again.
The messages where resolution by the user is possible are listed below. All other error messages
are listed in the service manual.
NOTE
Contact the manufacturer or specialist dealer if a message not described here is
displayed.
Possible cause
No mains voltage
The mains cable on the outlet or
device is not connected correctly
or at all
Device not switched on
Defective mains fuse
Internal power supply defective
Defective accessory
Defective device
"Auto-Start" has to be
acknowledged when loading a
program.
The button continues flashing
after it is acknowledged.
Even though the accessory was
removed, the button continues
flashing.
See listing below
Error rectification
Check the mains supply
Check the mains cable
connection
Switch on the device
Device defect, service required
Device defect, service required
Replace the accessory
Turn the device off and on again.
Notify Service of any error
message that is displayed.
Press the Auto-Start button. No
error.
Disconnect and check the
connected accessory.
Accessory is in contact with
tissue or defective.
Operating the CURIS
®
continue in the STANDARD
operating mode, then contact
Service.
See listing below
can
37