4.6.3 Example of a Calibration Certificate
Certificate No.:
Serial No. Rotor:
Anemometer No.:
Customer:
Inventory No.:
Total Flow Rate (L/min)
Non-Relevant Flow Rate (L/min)
Agar Strip Lot No.
Results refer to the Merck Standard Operating Procedures SOP-MB:PG-108, SOP-MB:PG-109 and SOP-MB:PG-232. They are traceable to the
Standards of the National Institute of Standards of Germany (Physikalisch Technische Bundesanstalt [PTB], Braunschweig, Germany)
Relevant Rotor Flow Rate [100 L/min ± 20%]:
Calibration Factor as received:
Running Time/300 L [correction factor: 0%]:
Calculated Calibration Factor (Cf):
Selected Calibration Factor [by software]:
Running Time/300 L [after adjustment]:
Verification of the Calibration Factor (within ± 2%)
Temperature:
Conform:
Yes / No
Date:
Customer notes (to be filled by the client):
Date of next calibration:
Certificate of Calibration
RCS High Flow Touch
Agar Strip 1
(delete as applicable)
Air Sampler
RCS High Flow
Serial No. Instrument:
Serial No. Protection Cap:
Anemometer Calibration Expiry:
Agar Strip 2
Setting:
Rel. Humidity:
Inspector:
Signature:
RCS Isolator
Agar Strip 3
Average
L/min
%
Re-tested
%
s
%
%
s
%
85