1 Introduction
1.5

Technical Data and Specifications

Collection principle
Operation
Range of measurement
Flow rate
Instrument deviation
Sample volume
Rotor speed
Tolerable axial force
on rotor shaft
Environmental
conditions
Electrical supply
Size (H × B × T), weight
Overvoltage category
Pollution
Automatic power off
Material
Interface
Validation
Calibration
User defined functions
58
(continued)
Centrifugal impaction according to Reuter (Reuter Centrifugal Sampler, RCS
Portable device with color touchscreen. Battery operation with integrated Lithiumion battery
1-2000 L (recommended: 10-1500 L)
100 L/min (1000 L in 10 min)
± 5 %
7 pre-set sample volumes: 10, 20, 50, 100, 200, 500 and 1000 L
3 user defined sample volumes from 1 to 2000 L
8200 rpm
30 Newton
Temperature: 5-40 ˚C, relative humidity: 20-80% (non-condensing)
Height: < 3000 M (NN)
• Rechargable Lithium-ion battery
• Power Supply (conversion direct current 24 V, power consumption 72 W)
• Docking Station (optional)
300 × 130 × 110 mm, approx. 1500 g (3.3 Ibs)
II
2
Within 8 minutes after the last action.
Housing: Lexan polycarbonate (PC); Sampling head (autoclave-able): Stainless steel/anodized aluminum/
APEC polycarbonate
Serial RS232 interface, USB adapter, standard tripod thread
According to ISO 14698 (with corresponding agar strips)
Automated calibration (HYCON
• Individual sample volumes
• Delay time
®
• Interval sampling (RCS
Management Software)
• QA Level management (RCS
• Language, local time
®
• Date (RCS
Management Software)
• Rotor selection and rotor management
• Optional software: HYCON-ID, CalibSo
®
Anemometer, CalibSo Software), calibration reminder
®