1.6
Electrical Supply
®
The RCS
High Flow Touch Microbial Air Sampler is
equipped with an integrated Lithium-ion battery to
allow for flexible operation independent from
mains.
Power Supply and a power cord are contained in
the RCS
®
High Flow Touch supply kit. Optionally,
a docking station (Art. No. 1.44256.0001) is avail-
able allowing for convenient recharging after each
use. Recharging the battery can be performed at
any time independent from the charging status.
NOTE
• Upon delivery the integrated battery of the
RCS
®
High Flow Touch Microbial Air Sampler
may be partially charged. It is recommended
to fully charge the battery prior to first use.
• Throughout the first charging cycles the total
capacity of the battery may slightly increase.
• Charging of the battery with power supply or
docking station can be performed at any time.
It is recommended to avoid complete discharg
ing of the battery on a regular basis.
• The minimal remaining battery capacity to start
the air sampling is set to two measurements
as per selected volume. This is a safety feature.
• Dependent on age and the frequency of usage
the total battery capacity will be reduced. This is
a technical feature of Lithiumion batteries.
It is recommended to perform at least an annual
diagnosis of the total battery capacity.
• It is recommended to cover the electrical sockets
with the corresponding protection caps upon
battery driven usage, or cleaning cycles.
59