Controller Screen
Display
9.
10.
11.
12.
®
eFlow
rapid nebuliser system – Questions about your device? Contact your service partner
Display Symbol/
Condition
Controller Sound
Flashing
• The device has reached
alternating signal
20-minute maximum time
tone
and will shut off.
Flashing
• Battery is empty, no
alternating signal
nebulisation is possible.
tone
Flashing
• Battery pack is defective.
alternating signal
tone
Flashing
• Battery pack is overheated. • Wait until the device cools
alternating signal
tone
EN
K
Troubleshooting
Action Required
• Check to see if medication
remains in the medication
reservoir (see Notes, Step I9).
• If medication remains, restart
the controller.
• After treatment is complete,
clean the aerosol head.
• Connect power supply to take /
resume treatment and keep
power supply plugged
afterwards to charge battery.
• Switch to power supply to
continue treatment.
• Contact your service partner.
down or switch to power
supply to continue treatment.