N Technical Data
Weight: controller and connection cord
Dimensions (LxWxH): controller (approx.)
Weight: nebuliser handset
Dimensions (LxWxH): nebuliser handset (approx.)
PARI power supply
(REF 078B7116 for international power supply)
Sound pressure level (nebuliser handset)
®
eFlow
rapid Performance Characteristics
Min. fill volume
Max. fill volume
a)
MMAD
(Mass Median Aerodynamic Diameter)
a)
GSD
(Geometric Standard Deviation)
a)
Respirable fraction
b)
Aerosol output
b)
Aerosol output rate
Aerosol data according to ISO 27427; nebulised medium: Salbutamol 2.5 ml. Measured values may vary depending on the medication used.
Recommended fill volume: see package leaflet for the medication
a) Measurement with Next Generation Pharmaceutical Impactor (NGI) at 23 °C and 50% relative humidity. Inspiratory flow: 15 l/min.
b) Measurement with breathing simulator at 23 °C and 50% relative humidity. 500 ml breathing volume, breathing frequency 15 cycles/minute, sinusoidal
breathing pattern, inhalation/exhalation ratio 1:1 (with adults, may vary in children).
For performance characteristics for a nebuliser handset specifically approved for and/or supplied with a drug refer
to the information about the drug.
®
eFlow
rapid nebuliser system – Questions about your device? Contact your service partner
EN
N
Technical Data
210 g / 8 oz
16 x 7 x 2 cm / 6.3 x 2.8 x 0.8 in
55 g / 1.9 oz
14.5 x 5.0 x 6.3 cm / 5.7 x 2 x 2.5 in
Input: 100 - 240 V, 50-60 Hz
Output: 5 V
Total power consumption < 3.5 Watt
< 33 dB (A)
< 5 μm
< 2 μm
2 - 5 μm
en
2.0 ml
6.0 ml
4.6 μm
1.8
55%
6.5%
48.5%
0.62 g
0.5 g/min
35