F
Disinfecting Your Nebuliser Handset
Step F5
Air-dry parts on a clean, lint-free towel.
Step F6
After the parts are completely dry, store them in a
clean bag or cover them with a clean, lint-free towel.
Reassemble just before treatment.
Method 2: Steam Disinfection
Steam disinfection can be performed with a commercially
available baby bottle disinfector with an operating time of at least
6 minutes.
Step F7
Follow the directions for the baby bottle disinfector
and use the full disinfection cycle with the correct
volume of water described in the directions.
Step F8
Remove the parts after the disinfector has cooled
and shake the parts to remove excess water.
Step F9
Air-dry parts on a clean, lint-free towel.
Step F10
After the parts are completely dry, store them in a
clean bag or cover them with a clean, lint-free towel.
Reassemble just before treatment.