Cleaning Of Materials; Disinfection Of The Product; Maintenance - Interco Simply Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für Simply:
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

Powder-coated aluminum
interco undercarriages and some add-on parts are
made of aluminum and feature a resistant plastic coat-
ing Powder varnishes are used for coating Powder
varnishes are solvent-free coating materials, com-
posed of different resin systems, pigments as well as
special additives

11.1.2. Cleaning of materials

Powder-coated aluminum
Clean undercarriages and attachments made of pow-
der-coated aluminum with mild soapy water and a soft
sponge Do not use aggressive or abrasive agents
These can damage the surface of the coating
Powder coated materials can be maintained either with
lukewarm, clear water or with the addition of neutral
to slightly alkaline cleaning agents� After the cleaning
process, rinse with clear water�
Accessory components made of metal, plastic, wood
Please clean accessory components with a conven-
tional mild household soap solution

11.1.3. Disinfection of the product

Please use a water-based agent for disinfection of
the product, eg Terralin, Quartamon Med or Sagro-
tan Original concentrate Please observe the manu-
facturer's instructions for use when doing so When
disinfecting the fabrics, also follow the instructions
of the fabric manufacturer regarding the use of brand
name disinfectants
Rinse off all detergents and disinfectants thoroughly
and leave the cleaned object to dry completely before
use If this is neglected, this can lead to chemical reac-
tions making the material brittle or lead to skin irrita-
tions of the patient
Important!
No disinfectants for synthetic leather/skai
Synthetic leather covers/skai covers
may not be cleaned with disinfecting agents, as
the material gets porous over time and cracks
Checking / components
Visual inspection of the seating system and the attached components and
accessories for visible damage
Check of connection between undercarriage and seating system
Visual inspection of all fastening elements
Visual inspection of all accessories / options
Retightening of all fastening elements
Retightening of all screw connections of the accessories / options
50
MAINTENANCE PL AN
Note!
Wipe disinfection
If wipe disinfection is necessary, use a
disinfectant that is suitable for surface
disinfection according to the DGH disinfectant list

11.2. Maintenance

Your interco product is checked by trained and quali-
fied employees for proper functioning and freedom
from defects before being handed over to the user
To ensure safe use of the product over its entire
service life (according to the manufacturer from at
least 2 years to 8 years maximum) and to ensure a
perfect technical condition, it must be maintained
and inspected regularly according to the following
maintenance (every 6 months) and service (every 12
months) plans
Note!
Changing settings
The changing of settings (other than
those described in Section 7) is only
permitted for qualified and trained personnel!
Incorrect settings by the user, operator or laypersons
bear the risk of injuries or crushing, which in a worst
case scenario could lead to the patient's death!
During normal use, you should regularly check the
interco product for damage and the perfect working
condition of all components
Important!
Original spare parts
Only original parts may be used for
service and maintenance work
Daily
Weekly
x
x
x
x
x