5
Commissioning
Pos: 21.2 /Standard-Layout/Hinweise/Wichtig/Wichtig - Hinweis auf SM EX-Sicherheitshinweise FM/CSA (OI/CI) @ 31\mod_1245414843234_3101.doc @ 280321 @ @ 1
Important
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas (Applies to FM / CSA only).
As a result, it is crucial that the specifications and data it lists are also observed.
Pos: 21.3 /Standard-Layout/Überschriften/1.1/1-spaltig/A - C/Allgemeine Informationen @ 7\mod_1166522996531_3101.doc @ 53066 @ 2 @ 1
5.1
General information
Pos: 21.4 /Standard-Layout/Inbetriebnahme/Durchfluss/FCM2000/Kontrollen vor Inbetriebnahme/Allgemeine Hinweise (CI) @ 26\mod_1237371759203_3101.doc @ 248283 @ @ 1
Inspection prior to switching on supply power
The following points must be checked before commissioning the device:
•
The assignment of the flowmeter sensor to the transmitter must be correct.
•
The wiring must be correct according to the electrical connections.
•
The flowmeter sensor must be correctly grounded.
•
The external data memory module (FRAM) must have the same serial number as the
flowmeter sensor.
•
The external data memory module (FRAM) must be inserted at the correct location (see the
section titled "Transmitter replacement" in the operating instructions).
•
The ambient conditions must meet the specifications.
•
The supply power must match the information on the name plate.
Inspection after switching on supply power
The following points must be checked after commissioning the device:
•
The parameter configuration must correspond to the operating conditions.
•
The system zero adjustment must have been made.
General information
•
If the flow direction indicated on the display is incorrect, it could mean that the signal lead
connections have been accidentally reversed.
•
The locations and the sizes of the fuses can be found in the section titled "Replaceable parts
list" in the operating instructions.