Pos: 21.1 /Standard-Layout/Überschriften/1/G - I/Inbetriebnahme @ 0\mod_1131980668835_3101.doc @ 3150 @ 1 @ 1
5
Commissioning
Pos: 21.2 /Standard-Layout/Hinweise/Wichtig/Wichtig - Hinweis auf SM EX-Sicherheitshinweise FM/CSA (OI/CI) @ 31\mod_1245414843234_3101.doc @ 280321 @ @ 1
Important
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas (Applies to FM / CSA only).
As a result, it is crucial that the specifications and data it lists are also observed.
Pos: 21.3 /Standard-Layout/Überschriften/1.1/1-spaltig/A - C/Allgemeine Informationen @ 7\mod_1166522996531_3101.doc @ 53066 @ 2 @ 1
5.1
General information
Pos: 21.4 /Standard-Layout/Inbetriebnahme/Durchfluss/FCM2000/Kontrollen vor Inbetriebnahme/Allgemeine Hinweise (CI) @ 26\mod_1237371759203_3101.doc @ 248283 @ @ 1
Inspection prior to switching on supply power
The following points must be checked before commissioning the device:
The assignment of the flowmeter sensor to the transmitter must be correct.
The wiring must be correct according to the electrical connections.
The flowmeter sensor must be correctly grounded.
The external data memory module (FRAM) must have the same serial number as the
flowmeter sensor.
The external data memory module (FRAM) must be inserted at the correct location (see the
section titled "Transmitter replacement" in the operating instructions).
The ambient conditions must meet the specifications.
The supply power must match the information on the name plate.
Inspection after switching on supply power
The following points must be checked after commissioning the device:
The parameter configuration must correspond to the operating conditions.
The system zero adjustment must have been made.
General information
If the flow direction indicated on the display is incorrect, it could mean that the signal lead
connections have been accidentally reversed.
The locations and the sizes of the fuses can be found in the section titled "Replaceable parts
list" in the operating instructions.
Pos: 21.5 /Standard-Layout/======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ @ 1
CI/FCM2000-EN
CoriolisMaster FCM2000
Commissioning
EN - 23