Herunterladen Inhalt Inhalt Diese Seite drucken

Electrical Connection - HBM C16i Montageanleitung

Vorschau ausblenden Andere Handbücher für C16i:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Total
Number
supply
of load
current
cells
1
110 mA
2
165 mA
3
220 mA
4
275 mA
6
385 mA
8
495 mA
10
615 mA
12
715 mA
16
935 mA
Load cell cable
length (max. 40 m)
Fig. 5.2
Connection diagram when C16i4 load cells are supplied via the digital
display
C16i
between the display and junction box
Nominal voltage U
0.14
0.25
2
mm
mm
101 m
180 m
67 m
120 m
50 m
90 m
40 m
72 m
29 m
51 m
22 m
40 m
18 m
33 m
16 m
28 m
12 m
21 m
Nominal voltage U
Max. cable length L
Junction box VKD2R-8
30 mA (bus termination)
C16i4 - 1
C16i4 - 2
A05516_01_Y00_01 HBM: public
Max. cable length L
12 VDC
b
Wire cross-section
2
0.5 mm
2
359 m
239 m
179 m
144 m
103 m
80 m
65 m
55 m
42 m
Digital display
b
RS485 + power supply
Total supply current
C16i4 - 3

Electrical connection

Nominal voltage U
0.14
0.25
2
2
mm
mm
212 m
379 m
141 m
253 m
106 m
190 m
85 m
152 m
61 m
108 m
47 m
84 m
39 m
69 m
33 m
58 m
25 m
45 m
Total
...
C16i4 - n
15 VDC
b
2
0.5 mm
500 m
500 m
378 m
302 m
216 m
168 m
137 m
116 m
89 m
19