the batteries are almost flat; the batteries must be re-
•
placed.
In the above cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk. Re-insert the batteries if neces-
sary, or else replace them.
8. Cleaning and storing the instrument
Clean your blood pressure computer carefully only with a
•
slightly moistened cloth.
Do not use detergents or solvents.
•
On no account must you immerse the computer in water,
•
otherwise liquid can enter it and cause demage.
When storing the device, make sure that no heavy ob-
•
jects are placed on top of it. Remove the batteries. The
cuff tube should not have any sharp kinks.
9. specifications
Model No.
Method of Measurement Oscillometric, non-invasive blood
Range of Measurement Pressure 0 – 300 mmHg,
Accuracy of the
pressure reading
BM 34
pressure measurement on the
upper arm
systolic 30 – 280 mmHg,
diastolic 30 – 280 mmHg,
pulse 40 – 180 beats/minute
systolic ± 3 mmHg / diastolic
± 3 mmHg / pulse ± 5 % of the
displayed value
Measuring uncertainty
Sets of Memory
Unit Dimension
Unit Weight
Cuff Size
Operating Temperature +5 °C to +40 °C, 40 – 85 % RH
Storage-Temperature
Power Supply
Battery Life
Accessories
Classification
Key to symbols:
These specification are subject to change without notice
for purpose of improvement.
This unit is in line with European Standard EN60601-1-2
•
and is subject to particular precautions with regard to
electromagnetic compatibility (EMC). Please note that
portable and mobile HF communication systems may
16
max. permissible standard devia-
tion according to clinical testing:
systolic: 8 mmHg /
diastolic: 8 mmHg
1 x 60 memory places
L137 x W97 x H55 mm
Approx. 260 g
Wrist circumference approx.
22 – 30 cm
-20 °C to +55 °C, 10 – 95 % RH
4 x 1,5 V AA batteries (alkaline
type LR6)
For approx. 180 measurements,
depending on blood pressure
level or inflation pressure
Instruction manu al,
4 pcs "AA" battery
Application part Type B
Important! Read the operating
instruction!
Application part Type B