Range of blood pressure values
Grade 3: Severe hypertension
Grade 2: Moderate hypertension
Grade 1: Mild hypertension
High-normal
Normal
Optimal
Source: WHO, 1999
The bar graph in the display and the scale on the unit
indicate the range of the blood pressure which has been
recorded.
If the values for systolic and diastolic pressure are in two
different WHO ranges (e.g. systolic in the high-normal
range and diastolic pressure in the normal range) the
graphic WHO classification on the unit indicates the higher
range (high-normal in the example described).

6. saving, retrieving and deleting results

The results of each successful measurement are stored
together with date and time. With more than 60 items of
measured data, the earliest items of data measured are
lost.
In order to retrieve measurement results again, press the
memory button „MEM". The instrument first displays the
number of memory spaces occupied and then automati-
cally switches to the last saved test result.
Systolic
Diastolic
(in mmHg)
(in mmHg)
> = 180
> = 110
160 – 179
100 – 109
140 – 159
90 – 99
130 – 139
85 – 89
120 – 129
80 – 84
< 120
< 80
Pressing the memory button „MEM" again outputs the
further measurement values with date and time.
You can clear the memory by holding down the memory
button „MEM" for 3 seconds.

7. error messages/troubleshooting

In case of faults, the „Er" message appears in the display.
Error messages can occur when
inflation takes longer than 3 minutes (Er6 appears in the
display),
the cuff is too tight or too loose (Er3 or Er4 appears in the
display),
the blood-pressure values are unusually high or low (Er1
or Er2 appears in the display),
the inflation pressure is higher than 300 mmHg (Er5 ap-
pears in the display),
there is a system or device error (Er0, Er7, Er8, Er9, ErA
or ErB appears in the display),
15
Measure
Seek medical advice
Seek medical advice
Have it checked regularly by doctor
Have it checked regularly by doctor
Check it yourself
Check it yourself