X-ray Exposure
4.3
Leave the X-ray room and close the door.
1.
Press down the Exposure Switch and hold it until image acquisition ends. The
2.
LED on the Exposure Switch turns yellow during the X-ray emission.
Release the Exposure Switch when you see the "Image capturing is completed"
3.
message on the screen.
Green X 18
TM
(Model: PHT-75CHS) User Manual
4. Acquiring CEPH Images (Optional)
▪
Release the Exposure Switch to stop X-ray emission
when an emergency occurs in the X-ray room during the
imaging.
▪
The operator shall always observe the X-ray safety
regulations applicable to his/her area during this
equipment's operation.
▪
The operator must always keep vocal/visual contact with
the patient during the image acquisition process.
▪
Do not operate the PC during exposure. Doing so may
cause the system to malfunction.
▪
To acquire the most optimized image, ask the patient:
▪
Keep his or her eyes closed
▪
Hold his or her breath and swallowing.
▪
Stay in his or her first position until the temple support is
open.
Yellow: X-ray On
During X-ray exposure, the status appears as follows.
▪
The LED light of the Exposure Switch turns yellow.
▪
The LED light on the top of the equipment turns yellow.
▪
An alert sound comes out to indicate that X-ray emission
is currently underway.
▪
On Console Software, the radiation mark turns green.
53