Fit the ear rods along the patient's ear canals to keep the patient's head in
6.
position. The patient's Frankfurt Line must be parallel to the floor.
Ask the patient to tilt his or her neck back to 30˚ - 40˚ as shown in the image
7.
below.
Ask the patient to swallow saliva and remain still until the X-ray exposure is
8.
complete.
Click the READY button after the patient has been properly positioned. No X-ray
9.
will be emitted at this point.
10. Go to 4.3 X-ray Exposure and follow the procedure for the X-ray exposure.
Green X 18
TM
(Model: PHT-75CHS) User Manual
4. Acquiring CEPH Images (Optional)
51