3. Acquiring PANO Images
X-ray Exposure
3.3
Leave the X-ray room and close the door.
1.
Press down the Exposure Switch and hold it until image acquisition ends. The
2.
LED on the Exposure Switch turns yellow during the X-ray emission.
Release the Exposure Switch when you see the "Image capturing is completed"
3.
message on the screen.
36
Release the Exposure Switch to stop X-ray emission
when an emergency occurs in the X-ray room during the
imaging.
The operator shall always observe the X-ray safety
regulations applicable to his/her area during this
equipment's operation.
The operator must always keep vocal/visual contact with
the patient during the image acquisition process.
Do not operate the PC during exposure. Doing so may
cause the system to malfunction.
To acquire the most optimized image, ask the patient:
Keep his or her eyes closed
Hold his or her breath and swallowing.
Stay in his or her first position until the temple support is
open.
Yellow: X-ray On
During X-ray exposure, the status appears as follows.
The LED light of the Exposure Switch turns yellow.
The LED light on the top of the equipment turns yellow.
An alert sound comes out to indicate that X-ray emission
is currently underway.
On Console Software, the radiation mark turns green.
Green X 18