3. Start the sterilization using the following parameters:
Conditioning Parameters:
Temperature: 55°C (131°F)
Humidity: 70% relative humidity
Conditioning dwell time: 30 minutes
Sterilization Parameters:
Sterilant: 100% ethylene oxide
Temperature: 55°C (131°F)
Gas concentration: 735±30 mg/l
Exposure Time: 180 minutes
Aeration Parameters
Time: At least 12 hours
Temperature: 55°C (131°F)
8.9.4 Chemical low-temperature sterilization with peracetic acid –
®
STERIS
CAUTION
Processing in the STERIS System 1E
devices cannot maintain sterility.
For sterilization with Steris
observed.
Use protective cap 8403 YZ for sterilization with Steris
Place the video laryngoscope in the appropriate tray.
Place the tray correctly into the SS1E and start the system.
®
Instruction For Use • C-MAC
®

System 1E

®
System 1E
Video Laryngoscope 8403xxx and Connection Cable 8403X • 96076019D • V4.3
Reprocessing USA and Canada
®
must occur immediately prior to use, since wet
®
, country-specific laws and regulations must be
®
System 1E
®
(see figure).