Videolaryngoskop
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

3. While immersed, use a soft bristle brush (Art. no.: 27652/3) to brush all the surfaces of the
device. Brush for a minimum of 1 minute.
4. Use a low-lint wipe to wipe the entire length of the connecting cable.
5. After the immersion and brushing period, remove the device from the cleaning solution and
rinse using a commercial overhead sprayer nozzle with cold water. Rinse for a minimum of
two minutes.
6. Dry the device and the contacts with a soft, lint-free cloth or dry, oil-free compressed air
(<5 psi).

8.7 Assembly, inspection and care

1. Check the product visually for purity, completeness, damage and dryness.
2. In case of soiling or residue, post-clean the product manually and carry out a complete
cleaning and disinfection process.
3. Sort out damaged and corroded products.
4. Assemble disassembled products.
5. Perform a functional check.
6. For care of the product, refer to the articles in the "Care, Sterilization and Storage
Technology" catalog.

8.8 Packaging

Only use standardized and approved packaging materials and packaging systems (EN 868
Part 2 – 10, EN ISO 11607 Part 1 + 2, DIN 58953).

8.9 Sterilization

The following sterilization method has been validated and approved for this product by KARL
STORZ:
8.9.1 Hydrogen peroxide (H₂O₂) sterilization - ASP STERRAD
WARNING
Risk of infection!
The use of STERRAD
lead to sufficient sterilization.
Sterilize only products that are not oiled and not greased.
Sterilize products in disassembled state.
WARNING
Different STERRAD
Please note that materials and lumen dimensions determine what products can be sterilized
using the various STERRAD
  ATTENTION  
Follow the instructions manual for the sterilizer!
The "STERRAD
can be sterilized using the different STERRAD
The following STERRAD
medical device by KARL STORZ:
®
Instruction For Use • C-MAC
®
processes on lubricated, oiled or pre-assembled products does not
®
sterilization systems
®
sterilization systems.
®
Sterility Guide" can be used to verify whether the relevant medical device
®
sterilization methods have been validated and approved for this
Video Laryngoscope 8403xxx and Connection Cable 8403X • 96076019D • V4.3
Reprocessing USA and Canada
®
devices.
®
83