5 Installation and commissioning

5.1 Visual and functional test

Visual inspection
en
The medical device must be inspected for completeness, damage and integrity before and
after every use. Tools such as a magnifying glass may be necessary to carry out the
inspections.
Missing components, surface changes or other damage to the medical device limit its life span
and may mean that the device can no longer be used for its intended purpose.
Check the instrument for:
Instruction For Use • C-MAC
– Mechanical damage and changes, for example, sharp edges, burred edges, rough
surfaces, protruding and bent parts (examples: Fig. 1 and 2)
– Signs of rust or corrosion
– Damage to the coating, such as shown in Fig. 3
– Complete and intact seals (that are not cracked, porous or swollen) and joints
– Residues and/or moisture present at the interfaces
– All components present and correctly positioned
– Mobility of all mobile components, if necessary once the components have been
assembled in the case of instruments that can be dismantled
– Complete, intact and securely positioned telescope and lens systems, e.g. with a
magnifying glass (examples Fig. 4 and 5)
®