Trouble-shooting
Error in temperature display
"F0"
Handle is not in its support
"F1"
No water (water pressure < 0,5 bar
"F2"
-5V is missing (supply voltage on the
control board)
"F3"
Break of the safety NTC
"F4"
Safety switching does not react
(49°C comparator)
"F5"
Break of the regulating NTC
"F6"
Different switch settings in the hand-
le holder
"F7"
Temperature too high (>48°C),
display only in the stimulation and
irrigation mode but not in the thermal
cleaning procedure
"F8"
Short-circuit of the regulating NTC
"F9"
No heating performance (excess
temperature switch), def. semi-con-
ductor relay or regulator)
If, errors cannot be corrected inform the ATMOS service staff. Do not attempt to repair the unit yourself!
50
UK
Insert handle in its support to allow thermal process for
reducing bacterial count to start (jet connection must
to the back of the unit.
Replace handle (Art.No. 502.0963.0)
Check, whether the water supply delivers a pressure of at
least 2 bar (did you open the water-tap?).
Filter clogged?
Inform the service staff.
Inform the service staff.
Have the safety switching (temperature controller) checked
by the service staff.
Inform the service staff.
Handle must be correctly inserted in its support.
Inform the service staff.
Check, whether temperature setting is too high.
If necessary, adjust desired temperature to a value of <48°C
by means of the coding switches.
Inform the service staff.
Have temperature feeler of the regulating NTC checked by the
service staff.
Inform the service staff (internal excess temperature switch
might have released).