Setting the Cylinder Diameter

ProntoPress-20
Instruction Manual
The cylinder diameter has to be set, in order to have the right
values of force or pressure. When a cylinder diameter is changed,
the database values of force or pressure in the methods will
automatically change to maintain the conditions in the mounting
cylinder.
Note: The counter which warns the user to clean the rams after
200 mountings, will be reset when a new cylinder diameter is
selected. It is therefore important to clean away any old cured
resin beneath the lower ram when changing cylinder diameter.
Please see the instructions in the 'Maintenance' section.
mO
Press MENU UP/DOWN
mN
parameters.
D
CONFIGURATION
Cylinder Diameter:
Temperature Unit:
Force or Pressure:
D
e
Press ENTER
Menu.
D
CYLINDER DIAMETER
25 mm
30 mm
1 1/4"
1 1/2"
D
mO
Press MENU UP/DOWN
mN
for the cylinder, which is mounted.
D
e
Press ENTER
D
Esc
Press Esc to return to the Select Method Group Menu.
Setting the Temperature Unit, Force or Pressure Mode, Force
Unit, Pressure Unit and Countdown Mode: See the section
Advanced Operations.
mON
to select the Cylinder
L:25 mm
R: 50 mm
CELSIUS
FORCE
Unit: kN
e
to activate the Cylinder Diameter
40 mm
50 mm
mON
to select the diameter
e
to accept the diameter.