Instructions
Intended Use
This device is intended for medical purposes to illuminate body surfaces.
It is used for non-invasive visual examination of intact skin.
This battery-operated product is designed for external examination only
within professional healthcare facilities by medical professionals.
Check the correct operation of the device before use! Do not use the
device if there are visible signs of damage.
CAUTION: Do not look directly into the LED light. Patients must close
their eyes during examinations.
Caution: UV Light. Do not look directly at light.
CAUTION: This product contains magnets. Do not use in proximity of
pacemakers or other magnetically sensitive devices or items.
In case of a serious incident with the use of this device, notify 3Gen
immediately and, if required by local regulations, your national health
authority.
CAUTION: Do not use the device in fire or explosive risk area (e.g.
oxygen-rich environment).
WARNING: This product can expose you to chemicals including methy-
lene chloride and hexavalent chromium, which are known to the State of
California to cause cancer or reproductive toxicity. For more information
go to www.P65Warnings.ca.gov.
Electromagnetic Compatibility
This device complies with the EMC Emissions and Immunity level re-
quirements of the standard IEC 60601-1-2:2014. The emissions charac-
teristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment
(for which CISPR 11 class B is normally required), this equipment might
not offer adequate protection to radio frequency communication services,
and the user might need to take mitigation measures, such as relocating
or re-orienting the equipment.
WARNING: Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper oper-
ation. If such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
WARNING: Use of accessories other than those provided by the man-
ufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and
result in improper operation.
WARNING: Portable RF transmitters should be used no closer than 30
cm (12 inches) to any part of the device. Otherwise, degradation of the
performance of this equipment could result.
Lumio 2 is intended for skin examinations with magnification, polarized
white light, and a range of UV light wavelengths.
Hold Lumio about 75-150 mm (3-6") above the skin to be examined.
Press the left button (W) to power on the device in polarized white
light mode. Pressing the left button (W) again – while the device is on
– toggles between white light and the most recently-used UV lighting
mode. To choose between UV light wavelengths, tap the (UV) button
until the UV indicator (UVI) corresponding with the desired wavelength
illuminates. Note that the "Wood" mode is a blend of UV wavelengths that
approximates the spectrum of a classic Wood's lamp.
Choose from one of three brightness levels by pressing the brightness
button (BR). The selected brightness level corresponds with the number
of brightness indicators (BI) illuminated.
Look through the lens with your eye placed 100-150mm (4-6") above the
device. Move the device closer or farther away from the skin to obtain the
desired image focus and field of view.
To turn off the unit, push and hold any button about 1 second. In order to
save power, the device will turn off automatically after 3 minutes.
OptiClip™ Accessory
Lumio 2 includes two magnetic accessories designed to optically en-
hance the examination. In order to increase the contrast of fluorescent
features during UV examinations, magnetically attach the OptiClip (OC)
labeled "495nm LONG PASS" at the dual attachment points (AP). In
order to increase the main lens magnification by an additional 2.5x, use
the "+2.5x" clip. One OptiClip may be stored in the handle's recessed
OptiClip Storage (OS).
Battery and Charging
CAUTION: This device uses a special 3.7V 4000mAh lithium ion battery,
which can only be purchased from 3Gen or from an authorized 3Gen
dealer. Do not under any circumstances use another battery other than
the one designed for this unit.
This device is equipped with a four-level Charge Indicator (CI). When
the unit is turned on and fully charged, all four LEDs light up. When less
than 75% of battery life remains, three LEDs are illuminated. If more than
half the battery is drained, two LEDs are on, while a single LED indicates
that less than 25% of battery capacity remains. To charge your device,
connect the supplied USB-C to USB cable to the Charging Port (CP) and
any IEC 60950-1 compliant (5V) USB port. The Charge Indicator (CI)
pulses to indicate charging, with the number of pulsing LEDs correspond-
ing with charge level. When charging is complete, all four indicator LEDs
illuminate. After years of use, you may wish to re-place the battery, which
is only available directly from 3Gen or an authorized 3Gen dealer. To
change the battery, contact 3Gen or your local 3Gen dealer.
Troubleshooting
Please check www.dermlite.com for the most current troubleshooting
information. If your device requires servicing, visit www.dermlite.com/
service or contact your local 3Gen dealer.
Care and Maintenance
CAUTION: No modification of this equipment is allowed.
Your device is designed for trouble-free operation. Repairs shall be made
only by qualified service personnel.
Cleaning
The exterior of your device, except the optical parts, may be wiped clean
with isopropyl alcohol (70% vol.) prior to use on a patient. The lens
should be treated as high-quality photographic equipment and should be
cleaned with standard lens cleaning equipment and protected from harm-
ful chemicals. Do not use abrasive material on any part of the equipment
or immerse the device in liquid. Do not autoclave.
Warranty: 10 years for parts and labor. The battery is warranted for 1
year.
Disposal
This device contains electronics and a lithium-ion battery that must be
separated for disposal and may not be disposed of with general house-
hold waste. Please observe local disposal regulations.
The Lumio 2 set includes:
Lumio 2 scope (polarized white LEDs, 365nm LEDs, 385nm LEDs,
405nm LEDs, 100mm 2.3x primary lens), 2.5x add-on lens, 495nm add-
on bandpass filter, protective pouch, USB to USB-C charging cable
Technical Description
Visit www.dermlite.com/technical or contact your local 3Gen dealer.

ENGLISH