Herunterladen Inhalt Inhalt Diese Seite drucken
Vorschau ausblenden Andere Handbücher für HUD:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 7
Instructions
Intended Use
This device is intended for medical pur-
poses to illuminate body surfaces. It is
used for non-invasive visual examination
of intact skin.
Check the correct operation of the de-
vice before use! Do not use the device if
there are visible signs of damage.
CAUTION: Do not look directly into the
LED light. Patients must close their eyes
during examinations.
In case of a serious incident with the use
of this device, notify DermLite imme-
diately and, if required by local regula-
tions, your national health authority.
CAUTION: Do not use the device in fire
or explosive risk area (e.g. oxygen-rich
environment).
WARNING: This product can expose
you to chemicals including methylene
chloride and hexavalent chromium,
which are known to the State of Cali-
fornia to cause cancer or reproductive
toxicity. For more information go to www.
P65Warnings.ca.gov.
Electromagnetic Compatibility
This device complies with the EMC
Emissions and Immunity level require-
ments of the standard IEC 60601-1-
2:2014. The emission characteristics of
this equipment make it suitable for use in
professional healthcare environment as
well as residential environment (CISPR
11 Class B). This equipment offers ade-
quate protection to radio communication
service. In the rare event of interference
to the radio communication service, the
user might need to take mitigation mea-
sures, such as relocating or re-orienting
equipment.
WARNING: Use of this equipment adja-
cent to or stacked with other equipment
should be avoided because it could re-
sult in improper operation. If such use is
necessary, this equipment and the other
equipment should be observed to verify
that they are operating normally.
WARNING: Use of accessories other
than those provided by the manufac-
turer of this equipment could result in
increased electromagnetic emissions or
decreased electromagnetic immunity of
this equipment and result in improper
operation.
WARNING: Portable RF transmitters
should be used no closer than 30 cm (12
inches) to any part of the device. Other-
wise, degradation of the performance of
this equipment could result.
IMPORTANT: Before initial use, please
charge the internal lithium ion battery.
See "Battery and Charging" for details.
HÜD is a dermatoscope with polarized
light designed for smartphone photogra-
phy of skin lesions.
DermLite HÜD fits on virtually any smart-
phone or tablet. For best results, remove
your protective case from your smart-
phone or tablet. Using the clip, attach
the device so that it is centered over
your smartphone's main camera.
Photography
Once you have chosen a skin lesion
to photograph, lightly moisten it and its
surrounding skin with hand sanitizer or a
similar liquid. This increases the translu-
cency of the skin and improves the quali-
ty of the image. If there is a lot of hair,
attach the cap (C). With HÜD attached
to your smartphone, place it directly on
the skin and push the power button (P)
to turn on the light. Capture the image
with your smartphone.
Battery and Charging
To conserve energy, this device is
designed to automatically turn off after
3 minutes. The battery can be expected
to supply power for about 5 three-minute
sessions. If the device turns off prior to
3 minutes, it may need to be recharged.
To charge your device, connect the
supplied to USB cable to the Charging
Port (CP) and any IEC 60950-1 compli-
ant (5V) USB port. The Charge Indicator
(CI) pulses while charging and remains
illuminated when fully charged.
Troubleshooting
Please check www.dermlite.com for the
most current troubleshooting informa-
tion. If your device requires servicing,
visit www.dermlite.com/service or con-
tact your local DermLite dealer.
Care and Maintenance
CAUTION: No modification of this equip-
ment is allowed.
Your device is designed for trouble-free
operation. Repairs shall be made only by
qualified service personnel.
Cleaning
The exterior of your device, except the
optical parts, may be wiped clean with
isopropyl alcohol (70% vol.) prior to
use on a patient. The lens should be
treated as high-quality photographic
equipment and should be cleaned with
standard lens cleaning equipment and
protected from harmful chemicals. Do
not use abrasive material on any part of
the equipment or immerse the device in
liquid. Do not autoclave.
Warranty: 2 years for parts and labor.
The battery is warranted for 1 year.
Disposal
This device contains electronics and a
lithium-ion battery that must be separat-
ed for disposal and may not be disposed
of with general household waste. Please
observe local disposal regulations.
This set includes a DermLite dermato-
scope, 2 caps, a microfiber pouch, and
a USB to Micro USB cable.
Technical Description
Visit www.dermlite.com/technical or
contact your local DermLite dealer.

ENGLISH

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Hud2

Inhaltsverzeichnis