Instructions
Intended Use
This device is intended for medical purposes to illuminate
body surfaces. It is used for non-invasive visual examination
of intact skin.
Check the correct operation of the device before use! Do not
use the device if there are visible signs of damage.
CAUTION: Do not look directly into the LED light. Patients
must close their eyes during examinations.
In case of a serious incident with the use of this device,
notify DermLite immediately and, if required by local regula-
tions, your national health authority.
CAUTION: Do not use the device in fire or explosive risk
area (e.g. oxygen-rich environment).
WARNING: This product can expose you to chemicals
including methylene chloride and hexavalent chromium,
which are known to the State of California to cause cancer
or reproductive toxicity. For more information go to www.
P65Warnings.ca.gov.
Electromagnetic Compatibility
This device complies with the EMC Emissions and Immunity
level requirements of the standard IEC 60601-1-2:2014. The
emission characteristics of this equipment make it suitable
for use in professional healthcare environment as well as
residential environment (CISPR 11 Class B). This equip-
ment offers adequate protection to radio communication
service. In the rare event of interference to the radio com-
munication service, the user might need to take mitigation
measures, such as relocating or re-orienting equipment.
WARNING: Use of this equipment adjacent to or stacked
with other equipment should be avoided because it could
result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to
verify that they are operating normally.
WARNING: Use of accessories other than those provided
by the manufacturer of this equipment could result in in-
creased electromagnetic emissions or decreased electro-
magnetic immunity of this equipment and result in improper
operation.
WARNING: Portable RF transmitters should be used no
closer than 30 cm (12 inches) to any part of the device. Oth-
erwise, degradation of the performance of this equipment
could result.
IMPORTANT: Before initial use, please charge the internal
lithium ion battery. See "Battery and Charging" for details.
HÜD is a dermatoscope with polarized light designed for
smartphone photography of skin lesions.
DermLite HÜD fits on virtually any smartphone or tablet. For
best results, remove your protective case from your smart-
phone or tablet. Using the clip, attach the device so that it is
centered over your smartphone's main camera.
Photography
Once you have chosen a skin lesion to photograph, lightly
moisten it and its surrounding skin with hand sanitizer
or a similar liquid. This increases the translucency of the
skin and improves the quality of the image. If there is a
lot of hair, attach the cap (C). With HÜD attached to your
smartphone, place it directly on the skin and push the power
button (P) to turn on the light. Capture the image with your
smartphone.
Battery and Charging
To conserve energy, this device is designed to automatically
turn off after 3 minutes. The battery can be expected to sup-
ply power for about 5 three-minute sessions. If the device
turns off prior to 3 minutes, it may need to be recharged. To
charge your device, connect the supplied to USB cable to
the Charging Port (CP) and any IEC 60950-1 compliant (5V)
USB port. The Charge Indicator (CI) pulses while charging
and remains illuminated when fully charged.
Troubleshooting
Please check www.dermlite.com for the most current trou-
bleshooting information. If your device requires servicing,
visit www.dermlite.com/service or contact your local Derm-
Lite dealer.
Care and Maintenance
CAUTION: No modification of this equipment is allowed.
Your device is designed for trouble-free operation. Repairs
shall be made only by qualified service personnel.
Cleaning
The exterior of your device, except the optical parts, may be
wiped clean with isopropyl alcohol (70% vol.) prior to use on
a patient. The lens should be treated as high-quality pho-
tographic equipment and should be cleaned with standard
lens cleaning equipment and protected from harmful chemi-
cals. Do not use abrasive material on any part of the equip-
ment or immerse the device in liquid. Do not autoclave.
Warranty: 2 years for parts and labor. The battery is war-
ranted for 1 year.
Disposal
This device contains electronics and a lithium-ion bat-
tery that must be separated for disposal and may not be
disposed of with general household waste. Please observe
local disposal regulations.
This set includes a DermLite dermatoscope, 2 caps, a
microfiber pouch, and a USB to USB-C cable.
Technical Description
Visit www.dermlite.com/technical or contact your local
DermLite dealer.

ENGLISH