Instructions
Intended Use
This device is intended for medical purposes to illuminate body surfaces.
It is used for non-invasive visual examination of intact skin.
This battery-operated product is designed for external examination only
within professional healthcare facilities by medical professionals.
Check the correct operation of the device before use! Do not use the
device if there are visible signs of damage.
CAUTION: Do not look directly into the LED light. Patients must close
their eyes during examinations.
In case of a serious incident with the use of this device, notify 3Gen
immediately and, if required by local regulations, your national health
authority.
CAUTION: Do not use the device in fire or explosive risk area (e.g.
oxygen-rich environment).
WARNING: This product can expose you to chemicals including methy-
lene chloride and hexavalent chromium, which are known to the State of
California to cause cancer or reproductive toxicity. For more information
go to www.P65Warnings.ca.gov.
Electromagnetic Compatibility
This device complies with the EMC Emissions and Immunity level re-
quirements of the standard IEC 60601-1-2:2014. The emission character-
istics of this equipment make it suitable for use in professional healthcare
environment as well as residential environment (CISPR 11 Class B). This
equipment offers adequate protection to radio communication service.
In the rare event of interference to the radio communication service,
the user might need to take mitigation measures, such as relocating or
re-orienting equipment.
WARNING: Use of this equipment adjacent to or stacked with other
equipment should be avoided because it could result in improper oper-
ation. If such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
WARNING: Use of accessories other than those provided by the man-
ufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and
result in improper operation.
WARNING: Portable RF transmitters should be used no closer than 30
cm (12 inches) to any part of the device. Otherwise, degradation of the
performance of this equipment could result.
IMPORTANT: Before initial use, please charge the internal lithium ion
battery. See "Battery and Charging" for details.
DermLite Foto X is a polarized and non-polarized dermatoscope de-
signed for use with a wide range of cameras and mobile devices.
IceCap®
To support your infection control measures, snap one of the included
disposable IceCaps (IC) over the installed faceplate (FP). To remove and
dispose, simply pull it off.
Foto X may be attached to many cameras or smartphones via either a
magnetic or purely threaded connection. For use with MagnetiConnect®
adapters (sold separately), screw the Ring Adapter (RA) onto the 37 mm
thread (37). To create a threaded connection with your camera, remove
the Ring Adapter (RA) and thread the Foto X onto a compatible camera
or threaded camera adapter*.
Place the unit onto the skin, then push the Power button (P) to activate
the light. Tapping the power button toggles between polarized and
non-polarized light. Move between four brightness levels via the Bright-
ness Buttons (BB). To adjust focus, turn the Spacer (S) as needed;
aligning the Focus Marks (FM) creates the ideal focal distance for most
photography.
To turn off your DermLite, push and hold the Power button (P) for about
one second; the unit turns off automatically after 3 minutes.
*Note that many cameras require the use of an additional adapter made
by the camera manufacturer or a photography supply company.
Removing the Spacer
The Spacer (S) may be removed by turning it counter-clockwise past a
slight resistance. When reattaching, make sure its three small protrusions
match up with the Bayonet Slot (BS).
Removing the Faceplate
The Faceplate (FP) may simply be pulled off the Spacer (S) by grabbing
it at the fingernail notches.
Battery and Charging
CAUTION: This device uses a special 3.7V 360mAh (2pcs) lithium ion
battery, which can only be purchased from 3Gen or from an authorized
3Gen dealer. Do not under any circumstances use another battery other
than the one designed for this unit.
This device is equipped with a four-level Charge Indicator (CI). When
the unit is turned on and fully charged, all four LEDs light up. When less
than 75% of battery life remains, three LEDs are illuminated. If more than
half the battery is drained, two LEDs are on, while a single LED indicates
that less than 25% of battery capacity remains. To charge your device,
connect the supplied USB-C to USB cable to the Charging Port (CP) and
any IEC 60950-1 compliant (5V) USB port. The Charge Indicator (CI)
pulses to indicate charging, with the number of pulsing LEDs correspond-
ing with charge level. When charging is complete, all four indicator LEDs
illuminate. After years of use, you may wish to re-place the battery, which
is only available directly from 3Gen or an authorized 3Gen dealer. To
change the battery, contact 3Gen or your local 3Gen dealer.
Troubleshooting
Please check www.dermlite.com for the most current troubleshooting
information. If your device requires servicing, visit www.dermlite.com/
service or contact your local 3Gen dealer.
Care and Maintenance
CAUTION: No modification of this equipment is allowed.
Your device is designed for trouble-free operation. Repairs shall be made
only by qualified service personnel.
Cleaning
The exterior of your device, except the optical parts, may be wiped clean
with isopropyl alcohol (70% vol.) prior to use on a patient. The lens
should be treated as high-quality photographic equipment and should be
cleaned with standard lens cleaning equipment and protected from harm-
ful chemicals. Do not use abrasive material on any part of the equipment
or immerse the device in liquid. Do not autoclave.
Warranty: 2 years for parts and labor. The battery is warranted for 1 year.
Disposal
This device contains electronics and a lithium-ion battery that must be
separated for disposal and may not be disposed of with general house-
hold waste. Please observe local disposal regulations.
The DermLite Foto X set includes:
Foto X scope (24 polarized & 8 non-polarized LEDs, 30mm 8x lens),
USB-A to USB-C charging cable, 5 IceCaps.
Technical Description
Visit www.dermlite.com/technical or contact your local 3Gen dealer.