NOTE
Without a Telnet connection, the flashing LED indicators are displayed against
a background of the constant green LED. As soon as a Telnet connection has
been set up (even if established with CMD Assistant), the flashing LED
indicators are displayed against the backgorund of a constant blue LED.

3.3 Sensor connection

The charge amplifier CMD600 has a charge input (measurement channel) to
which piezoelectric sensors can be connected. All piezoelectric sensors
generating max. 600,000 pC during measurement can be used. IEPE sensors
cannot be operated (see chapter 11 "Tips regarding piezoelectric
measurement technology").

3.4 Ethernet interface

This interface can be used to integrate the CMD600 in a control or connect it
to a PC for parameterization.
The interface also permits measured values to be sent to a host processor −
via UDP streaming.
You can parameterize with the CMD Assistant software on the CD or with a
command set (see Chapter 7.2).
This CD also contains the Lab
can use to integrate the CMD600 into the National Instruments software.
Moreover, the CD comprises an example PLC which enables the CMD600 to
be integrated and operated in a Siemens−S7 control via TCP/IP.
Tip:
The latest Assistant and all other software tools are available at
www.hbm.com/support
A2781−2.1 en/de/fr
View software (driver and library) which you