3.8 Parameter sets

Two power failsafe parameter sets are saved in the CMD600 (flash)
Parameter sets can be selected with
• the CMD wizard
• digital input 1
• interface command (see operating manual "Command Set CMD600")
The switching time between parameter sets is:
5 ms without range switching
160 ms with range switching
If the internal measuring range of 6000 pC is exceeded, a reset is performed
automatically.
CAUTION
System parameters (for example IP/device address) are not overwritten
when the parameter set is changed.
An overview of system parameters can be found in the operating manual
"Command Set CMD600".
If the SensorTeach additional function is active, the parameter set cannot be
changed.
When the CMD600 is powered on, the last parameter set to be active is
loaded.
The first time the CMD600 is placed in operation, parameter set 1 is always
active.