Chapter 6. Troubleshooting

Problems
The pump stops.
Red ALARM LED lights up.
No cannula suction.
No shaver suction.
The roller pump heads do
not rotate.
Potential Causes
The HIGH PRESSURE safety
has activated.
The LOW PRESSURE safety
has activated: "Pres. L" blinks.
The filter is wet. (see Sections
3.2.A and 4.3)
There is no more irrigation fluid.
The cannula is obstructed.
The red clamp is closed.
The shaver blade is obstructed.
The Pinch Valve is
malfunctioning.
The power is off "O" or is on Stop. Check that the main power switch
36
Possible Solutions
Lower the joint pressure.
Press the Run/Stop pad to restart the
pump.
Verify that the luer-lock is tightly
closed.
Press the Fill Chamber to stop the
alarm, and press the Run/Stop pad to
restart the pump.
Replace the One Day Set Irrigation
Tube. If problem persists, contact
DePuy Mitek Customer Service.
Spike the bag and press the Fill
Chamber pad.
Reposition or clean the cannula.
Open red clamp.
Disassemble the blade and clean it.
Check the positioning of the suction
tubings.
Check the interface cable connection.
on the back of the pump is on the "I"
position, and that the Run/Stop pad
green light is on.