SUCTION roller pump.
8
Provides outflow of fluid.
9
Pinch Valve.
Moves automatically when the
shaver is in rotation, selecting the
appropriate suction tube.
Lavage mode pad.
10
Automatically starts a two-minute
sequence of pressure and flow
increases to clear blood and debris
from the field of view.
Pinch Valve pad.
11
This automatically opens the
pinch valve to insert or remove the
suction tubings.
12
SHAVER Speed Selection pads.
Select Low, Med or High pre-set
shaver speeds.
13
SHAVER Run/Stop pad.
Starts and stops the shaver mode
when using the FMS shaver.
14
SHAVER Suction setting.
Indicates the suction rate selected
through the shaver (Min, Low, Med,
High).
15
SHAVER Speed increase/
decrease pads.
Increases or decreases the shaver
rotation speed (rpm - rotation per
minute).
16
SHAVER handpiece connection.
Line up the red dots on the male
and female connectors to connect
the shaver handpiece.
Cannula suction setting.
17
Sets the flow rate through the
cannula (Min, Low, Med, High).
PUMP Run/Stop pad.
18
Turns the pump on and off. If the
pump is activated, PUMP appears
on the pressure display. Green LED
blinks when the pump is stopped.
Built-in pressure transducer
19
connection.
Measures the pressure in the
pressure chamber of the One Day
Set Irrigation Tube.
Fill Chamber pad.
20
Manually fills the chamber and
turns off the LOW PRESSURE
alarm.
PRESSURE adjustment pads.
21
Press the arrows to increase or
decrease the pressure level in
increments of 5. The initial setting is
50.
Auto-locking device.
22
This patented device allows the
disposable tubing to be properly
positioned and locked in place
around the two roller pumps. For
safety reasons, it has been color-
coded green for the IRRIGATION
and orange for the SUCTION
tubing.
15