Herunterladen Inhalt Inhalt Diese Seite drucken

Inspection Schedule; Annual Inspection - DePuy Mitek FMS DUO+ Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 152

Inspection Schedule

Weekly inspection:
DePuy Mitek recommends a weekly inspection of
all equipment. These inspections can be performed
on a less regular basis depending on the duration
and frequency of use.
- Perform a visual inspection of the device. If
damage is noted contact your DePuy Mitek
Sales Representative or call DePuy Mitek
Customer Service in the USA at +1-800-382-
4682, option 2 to speak to a bio-med.
- Inspect that the transparent safety covers (1)
and black hinges are in good working order.
- Inspect power cord and all cables for wear or
cuts.
- Ensure that the foot pedal and handpiece
connector pins are not missing or bent.
- Plug the power cord to the power connection.
- Switch the I/O switch (23) on "I".
- Turn the device ON by pressing the POWER (4)
pad.
- Press on all pads. An audible tone should be
heard for each pad pressed, except for Lavage,
(Pad 10) (Lavage is inactive until the pump is
primed.)

Annual inspection:

DePuy Mitek recommends that a bio-medical
engineer or authorized service personnel annually
check all equipment to assess its functionality and
technical safety.
Refer to Instruction Manual and Service Manual for
annual inspection procedures.
8