Herunterladen Inhalt Inhalt Diese Seite drucken

Patient Set Suction Tube Set-Up - DePuy Mitek FMS DUO+ Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 152
II. Patient Set Suction Tube Set-up.
Red clamp
(Y)
(Z)
This tube allows suction from the joint, either from a cannula (Y) or through a shaver (Z).
Do not close the red clamp. The red clamp should only be used to reduce or stop flow during surgery.
The scrub nurse passes the
waste extremity of the suction
tube to the non-sterile person
(circulating nurse).
Place and center the tube
around the roller pump.
ORANGE
Auto-locking device
The circulating nurse inserts
the waste end in a waste
collection bag (Ref. 281380).
23
To waste
collection
bag
Slide the orange part of the
tube into the auto-locking
device on the pump.